Health product recall

fabian Therapy evolution (2018-12-17)

Starting date:
December 17, 2018
Posting date:
December 28, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type I
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68730

Affected products

fabian Therapy evolution

Reason

A manufacturing defect in the front membrane panel overlay could cause the panel buttons to be electrically shorted, thus appearing to the system that the button is being depressed. When this happens, the system will continue to power cycle with alarming. Under the worst circumstances, if the ventilator ceases to deliver positive pressure breaths the patients the patient might experience transient systemic arterial hypercapnia and oxygen desaturation secondary to hypoventilation.

Affected products

fabian Therapy evolution

Lot or serial number

AT-01572 to AT-03025

Model or catalog number

7250.N

Companies
Manufacturer

Acutronic Medical Systems AG

Fabrik im Schiffli

Hirzel

8816

SWITZERLAND