fabian Therapy evolution (2018-12-17)
- Starting date:
- December 17, 2018
- Posting date:
- December 28, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type I
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-68730
Affected products
fabian Therapy evolution
Reason
A manufacturing defect in the front membrane panel overlay could cause the panel buttons to be electrically shorted, thus appearing to the system that the button is being depressed. When this happens, the system will continue to power cycle with alarming. Under the worst circumstances, if the ventilator ceases to deliver positive pressure breaths the patients the patient might experience transient systemic arterial hypercapnia and oxygen desaturation secondary to hypoventilation.
Affected products
fabian Therapy evolution
Lot or serial number
AT-01572 to AT-03025
Model or catalog number
7250.N
Companies
- Manufacturer
-
Acutronic Medical Systems AG
Fabrik im Schiffli
Hirzel
8816
SWITZERLAND