Fabian HFOi Ventilator
Brand(s)
Last updated
Summary
Product
Fabian HFOi Ventilator
Issue
Medical devices - Software defect
What to do
Contact the manufacturer if you require additional information.
Affected products
Affected Products |
Lot or Serial Number |
Model or catalog number |
|---|---|---|
Fabian HFOi Ventilator |
More than 10 numbers, contact manufacturer. |
113001, 7200.I |
Issue
Bias flow selection for NO system usage is missing from the user interface of Fabian HFOi Ventilators following incorrect selection of device configuration during a software update. This is caused by human error and there is no defect in the device or its software.
Recall start date: Mar 28, 2022
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies
Acutronic Medical Systems Ag
Fabrik Im Schiffli, Hirzel
Switzerland, 8816
Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type I
Identification number
RA-64066
Get notified
Receive emails about new and updated recall and safety alerts.