Health product recall

EZ1 Advanced XL (2018-07-25)

Starting date:
July 25, 2018
Posting date:
August 2, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67354

Affected products

EZ1 Advanced XL

Reason

QIAGEN was alerted by the supplier of an issue with a component on the printed circuit board.  The issue has the potential to trigger a heater malfunction. This could cause a delay in obtaining results from downstream applications.

Affected products

EZ1 Advanced XL

Lot or serial number

L16-L18 ending with 1801-2137

Model or catalog number

9001492

Companies
Manufacturer
QIAGEN GmbH
QIAGEN Strasse 1
Hilden
40724
GERMANY