Health product recall

Extracorporeal Tubing Set for Heart-Lung Machine - Perfusion Pack and Art.Cannula - Wire Reinforced, Straight Tip W/Ring, Pre-Mounted Connector (2018-05-07)

Starting date:
May 7, 2018
Posting date:
June 1, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66946

Affected products

A. Extracorporeal Tubing Set for Heart-Lung Machine - Perfusion Pack
B. Art Cannula -  Wire Reinforced, Straight Tip W/Ring, Pre-Mounted Connector

Reason

An arterial cannula used for open heart surgery is normally equipped with a vent plug, which facilitates the de-airing of the cannula. The vent plug of our arterial cannula consists of sintered polyethylene, therefore the structure of the vent plug is porous and gas-permeable. After the de-airing process is completed the customer has to separate the vent plug from the cannula to connect the extracorporeal circuit. In some rare cases it has come the vent plug broke during the separation from the arterial cannula.

Affected products

A. Extracorporeal Tubing Set for Heart-Lung Machine - Perfusion Pack

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

H 94500

Companies
Manufacturer
Maquet Cardiopulmonary GmbH
Kehler Str. 31
Rastatt
76437
GERMANY

B. Art Cannula -  Wire Reinforced, Straight Tip W/Ring, Pre-Mounted Connector

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

A20-0106
A20-0107
A22-0106
A22-0107
A24-0107

Companies
Manufacturer
Maquet Cardiopulmonary GmbH
Kehler Str. 31
Rastatt
76437
GERMANY