Extension Table Top
Brand(s)
Last updated
Summary
Product
Extension Table Top
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
Extension Table Top | 5 | 1140.20F0 |
Issue
For commercially released 1140.20f0 operating table tops it has been determined that a coincidence of 1. Lacking maintenance, 2. Not following the IFU in terms of lever orientation to prevent collision with the extension bars, and 3. Inattention to if at least one side is locked properly may lead to a loss of supporting functionality.
Recall start date: October 24, 2022
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies
Maquet GmbH
Kehler Strasse 31, Rastatt, Germany, 76437
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-72142
Get notified
Receive notifications for new and updated recalls and alerts by category.