Exoplan
Brand(s)
Last updated
Summary
Product
Exoplan
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalogue number |
---|---|---|
Exoplan | (01)4260521365019(10)A03B00E** 426052136EXOPLAN21A6 (01)4260521365002(10)A02B03E** (01)4260521365026(10)A03B01E** | 2.3 MATERA 3.0 GALWAY 3.1 RIJEKA |
Issue
During an update of data used in a library shipped with Exoplan or made available on the download portal of Exocad, it was found that the use of a specific combination of an implant, sleeve and surgical drill can result in a hole being drilled 1 mm deeper than expected.
Recall start date: October 12, 2023
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Exocad Gmbh |
Rosa-Parks-Str.2, Darmstadt, Germany, 64295 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74654
Get notified