EXCELSIOR AS TISSUE PROCESSOR (2016-01-22)
- Starting date:
- January 22, 2016
- Posting date:
- February 12, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-57106
Affected Products
A. EXCELSIOR AS TISSUE PROCESSOR
Reason
A software bug has been identified where 248.5513 days after being switched on the Thermo Scientific Excelsior as software switches from a positive to a negative count cycle.
Affected products
A. EXCELSIOR AS TISSUE PROCESSOR
Lot or serial number
Not applicable
Model or catalog number
A82300001