Health product recall

Evis Exera III Video System - Video System Center

Last updated

Summary

Product
Evis Exera III Video System - Video System Center
Issue
Medical devices - Device compatibility
What to do

Contact the manufacturer if you require additional information. 

Affected products

Affected products

Lot or serial number

Model or catalogue number

Evis Exera Iii Video System - Video System Center

7384649
7384742
7384689
7384743
7384740

CV-190

Issue

Olympus found that certain lots of the CV-190 do not start up properly and as a result, the image from the endoscope is not displayed when the operations listed below are performed because parts that deviated from the specification were assembled into the power supply unit.

Recall start date: September 12, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies
Olympus Medical Systems Corp.
2951 Ishikawa-Cho, Hachioji-Shi, Tokyo-To, Japan, 192-8507
Published by
Health Canada
Audience
General public
Healthcare
Industry
Recall class
Type II
Identification number
RA-74310

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe