Evis Exera III Video System - Video System Center
Brand(s)
Last updated
Summary
Product
Evis Exera III Video System - Video System Center
Issue
Medical devices - Device compatibility
What to do
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Evis Exera Iii Video System - Video System Center |
7384649 |
CV-190 |
Issue
Olympus found that certain lots of the CV-190 do not start up properly and as a result, the image from the endoscope is not displayed when the operations listed below are performed because parts that deviated from the specification were assembled into the power supply unit.
Recall start date: September 12, 2023
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies
| Olympus Medical Systems Corp. |
| 2951 Ishikawa-Cho, Hachioji-Shi, Tokyo-To, Japan, 192-8507 |
Published by
Health Canada
Audience
General public
Healthcare
Industry
Recall class
Type II
Identification number
RA-74310
Get notified