Health product recall

Evis Exera III Video System - Gastrointestinal Videoscope

Last updated

Summary

Product
Evis Exera III Video System - Gastrointestinal Videoscope
Issue
Medical devices - Labelling and packaging
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Evis Exera III Video System - Gastrointestinal Videoscope

More than 10 numbers, contact manufacturer.

GIF-1TH190

Issue

Olympus previously issued revised cleaning instructions for the GIF-1TH190 on October 11, 2023, to issue new updated channel drying steps. Olympus conducted user evaluation, through human factors validation, and determined further clarification is required to improve the instructions for easier understanding of each drying step. As a result, Olympus is issuing a revised IFU addendum that informs customers that the channel drying time has been extended by 12 minutes to purge water from the channels. This is accomplished by drying each channel for 3 minutes, in the sequence described in the addendum. Ensure all reprocessing personnel are completely knowledgeable and thoroughly trained on the new water purge requirements. Olympus has received 16 reportable complaints related to this issue.

Recall Start Date: October 18, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies

Olympus Medical Systems Corp.

2951 Ishikawa-Cho, Hachioji-Shi, Tokyo-To, Japan, 192-8507

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76429

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe