Health product recall

Everest Cutting Forceps

Brand(s)
Last updated

Summary

Product
Everest Cutting Forceps
Issue
Medical devices - Sterility
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number
Everest Cutting Forceps FR204443 FR176934 FR246906 FR305512 FR215026 FR141534 FR165250 3005

Issue

Olympus has received 15 complaints during the December 2018 - December 2023 timeframe from customers indicating fractures and breakages in packaging trays and Tyvek covers which may lead to a sterility breach. The root cause of the broken trays is related to a packaging change. Olympus has received no reports of adverse events related to the identified issue.

Recall Start Date: July 2, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies

Gyrus Acmi, Inc

9600 Louisiana Ave. North, Brooklyn Park, Minnesota, United States, 55445

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76047

Get notified

Receive emails about new and updated recall and safety alerts.