LivaNova has received reports of the Cockpit Graphical User Interface (GUI) of the Essenz HLM with SW HLM.1.5 activating a self-resetting safety feature, with the operation of the GUI being restored during use. During the reported events, the touchscreen display has temporarily gone dark (black/grey) and reset back to the home screen after approximately 10-15 seconds.