Essenta Dr Compact Digital Radiography System
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products | Lot or serial number | Model or catalogue number |
---|---|---|
Essenta Dr Compact Digital Radiography System | Not applicable. | 9890 010 88021 |
Issue
There is a potential cybersecurity issue with Essenta Dr Compact Systems, which could result in unauthorized disclosure or modification of patient data. This issue may occur if a firewall is not in place, the file transfer protocol (ftp) is utilized, and the following sequence of events occurs:
• unauthorized personnel gains physical access to customer facility
• customer facility does not have a firewall and/or network segregation to isolate medical devices, and
• the unauthorized person connects their device to the Essenta Dr Compact System network at the specific time data is transferred (data is not continuously transferred).
Under these conditions unauthorized access or modification of patient data is possible.
Recall start date: August 18 , 2023
Additional information
Details
Philips Medical Systems Dmc Gmbh |
Roentgenstrasse 24, Hamburg, Germany, 22335 |
Get notified
Receive notifications for new and updated recalls and alerts by category.