This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Esprit Ventilator (2014-12-08)

Starting date:
December 8, 2014
Posting date:
January 14, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-43289

Recalled Products

Esprit Ventilator

Reason

Esprit V1000 manufactured prior to December 31, 2004 may have the power supply fan mounting screws installed incorrectly, which could cause the ends of the screws to touch the power supply PCB-A and short to ground.

Affected products

Esprit Ventilator

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

V1000

Companies
Manufacturer

Respironics California, Inc.

2271 Cosmos Court

Carlsbad

92011

California

UNITED STATES