EPIQ Diagnostic Ultrasound System
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
EPIQ Diagnostic Ultrasound System |
All lots. |
EPIQ |
Issue
Philips EPIQ Ultrasound systems operating with software version 10.0 may present an image lag, where the user is expecting to see a real time image and the system delays the presentation of that image by several seconds. If present, this issue will be observed in the following scenario:
• The user begins to live scan and they do not see an image for several seconds (image is not immediately displayed).
• The image lag will continue to persist (image displayed will continue to be several seconds behind).
Recall Start Date: October 13, 2023
Additional information
Details
Philips Ultrasound Llc
22100 Bothell Everett Highway, Bothell, Washington, United States, 98021-8431
Get notified
Receive notifications for new and updated recalls and alerts by category.