Health product recall

Entroy Pool Lift (2018-03-16)

Starting date:
March 16, 2018
Posting date:
September 28, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67834

Affected products

Entroy Pool Lift

Reason

A potential for the chair or stretcher detachment from Entroy pool lifter is caused by a fact that caregiver may suppose that the pin has been paired with the seat/stretcher, but there could be a misalignment that might be unnoticed by the caregiver so the seat/stretcher does not get fully paired.  Instead, once the lift reaches a certain height, the seat/stretcher falls off the lift. To raise awareness and reduce risk a quick reference guide will be installed with the Entroy pool lift device.

Affected products

Entroy Pool Lift

Lot or serial number

221445

Model or catalog number

GAB1000-01-US

Companies
Manufacturer
ArjoHuntleigh AB
Hans Michelsensgatan 10
Malmo
21120
SWEDEN