Enterprise Imaging XERO Viewer
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Enterprise Imaging XERO Viewer |
All Versions |
8.X |
Issue
Agfa Healthcare has become aware of an issue related to orientation of 3D volumes using reference lines, active target localization, and series linking across multiple planes. A coding error has been identified in the initialization of pixel spacing in the XERO Viewer. This issue only applies to XERO Viewer. It does not apply to the XERO Viewer extended tools (Xtend) component or the enterprise imaging diagnostic desktop viewer.
Recall Start Date: September 10, 2024
Additional information
Details
Agfa Healthcare N.V.
Septestraat 27, Mortsel, Belgium, 2640
Get notified
Receive notifications for new and updated recalls and alerts by category.