Enteralite and Infinity Orange Pumps
- Starting date:
- November 8, 2012
- Posting date:
- December 10, 2012
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public (GP), Healthcare Professionals, Hospitals
- Identification number:
- RA-16275
Recalled Products
A. Enteralite and Infinity Orange Pumps
Reason
The recall concerns an error ER13 for Zevex's Moog Enteralite Feeding Pumps manufactured between the dates August 16, 2010 and April 15, 2011. The error is caused by defective diodes. Even if the diodes pass the manufacturing controls testing, there still is a possibility of pump failure. The recall is being initiated to ensure all pumps in field receive the correction.
Affected products
A. Enteralite and Infinity Orange Pumps
Lot or serial number
More than 100 numbers, contact manufacturer
Model or catalog number
INFKIT2
Companies
- Manufacturer
-
Zevx Inc., also trading as Moog Medical Devices Group