Health product recall

Engstrom Pro, Carescape, Engstrom Carestation

Last updated

Summary

Product
Engstrom Pro, Carescape, Engstrom Carestation
Issue
Medical devices - Electrical Defect
What to do

Contact the manufacturer if you require additional information.

Affected products

Affected Products Lot or Serial Number Model or catalog number
Engstrom Pro - Main Unit More than 10 numbers, contact manufacturer. 1505-9003-000
Carescape R860 More than 10 numbers, contact manufacturer. 1506-8600-000
Engstrom Carestation - Main Unit More than 10 numbers, contact manufacturer. 1505-9000-000

Issue

Carescape R860 ventilators, and Engström Carestation and Engström Pro ventilators with affected field replacement batteries - insufficient battery backup power resulting in premature shutdown of the ventilator when not connected to AC mains power supply.

Recall start date: Apr 26, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies

Datex-Ohmeda, Inc.

3030 Ohmeda Drive, Madison, Wisconsin

United States, 53707-7550

Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type II
Identification number
RA-64153

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe