Health product recall

Engstrom Pro, Carescape, Engstrom Carestation

Last updated

Summary

Product
Engstrom Pro, Carescape, Engstrom Carestation
Issue
Medical devices - Electrical defect
What to do

Contact the manufacturer if you require additional information.

Affected products

Affected Products

Lot or Serial Number

Model or catalog number

Engstrom Pro - Main Unit

More than 10 numbers, contact manufacturer.

1505-9003-000

Carescape R860

More than 10 numbers, contact manufacturer.

1506-8600-000

Engstrom Carestation - Main Unit

More than 10 numbers, contact manufacturer.

1505-9000-000

Issue

Carescape R860 ventilators, and Engström Carestation and Engström Pro ventilators with affected field replacement batteries - insufficient battery backup power resulting in premature shutdown of the ventilator when not connected to AC mains power supply.

Recall start date: Apr 26, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies

Datex-Ohmeda, Inc.

3030 Ohmeda Drive, Madison, Wisconsin

United States, 53707-7550

Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type II
Identification number
RA-64153