Health product recall

Endopath Echelon Vascular Reloads

Last updated

Summary

Product
Endopath Echelon Vascular Reloads
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Endopath Echelon Vascular Reloads

All lots distributed since 2024-05-30 and with expiry dates starting 2026-08-31

VASECR35

Issue

An increase in firing issues complaints resulting from premature sled movement lockout (PMSLO) in the field. PMSLO is a safety mechanism that prevents the device from firing with a spent reload or no reload. It can accidentally be triggered if the cartridge is bumped and the sled moves out of position. Investigation has revealed that the force to bump the sled from the home position in the newest cartridge body tool is statistically lower than the previous tool, although there are no dimensional non-conformances. The IFU describes the steps to reduce risk of harm when using the device on critical vessels (i.e. proximal and distal control) and steps to safely remove the device if it locks on tissue.

Recall start date: April 29, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies
Ethicon Endo-Surgery, LLC
475 Calle C, Guaynabo, Puerto Rico, United States, 00969
Published by
Health Canada
Audience
Healthcare
Recall class
Type I
Identification number
RA-77376

Get notified