Endoflip System
Brand(s)
Last updated
Summary
Product
Endoflip System
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Endoflip System |
All lots. |
DLEF-100 |
Endoflip System |
All lots. |
EF-100 |
Issue
The issue concerns a problem affecting EF and ES catheters used with EF-100 systems after 1-January-2026. Following this date will be incorrectly flagged as expired due to the system's 24-hour usage limit, rendering the catheter unavailable for clinical procedures.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Gastroenterology and urology
Companies
| Covidien LLC |
| 15 Hampshire Street, Mansfield, Massachusetts, United States, 02048 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type I
Recall date
Identification number
RA-82078