Health product recall

ENDOBON XENOGRAFTANULE GRS (2016-01-25)

Starting date:
January 25, 2016
Posting date:
February 8, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-57018

Affected products

ENDOBON XENOGRAFTANULE GRS

Reason

The cytotoxicity study of Endobon Xenograft Granules revealed severe (grade 4) toxicity after 36 months real-time ageing. This finding was also reported after accelerated ageing equivalent to 36 months real-time.

Affected products

ENDOBON XENOGRAFTANULE GRS

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number
  • ROX05
  • ROX10
  • ROX20
  • ROXLG20
  • ROXLG50
  • ROXLG80
Companies
Manufacturer

Biomet France Sarl

VALENCE

FRANCE