Endeavour Light System (2013-10-28)
- Starting date:
- October 28, 2013
- Posting date:
- November 29, 2013
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-36929
Recalled Products
Endeavour Light System
Reason
If user fails to secure the cable to the temple arm of his/her loupes with the spiral temple clip provided with the device, strain on the cable (cable pull forces in excess of 7 pounds) where it interfaces to the headlight may result in an electrical open that prevents power from reaching the headlight LED.
Affected products
Endeavour Light System
Lot or serial number
78100100 to 78113815
Model or catalog number
922003-X
Companies
- Manufacturer
-
Kerr Corporation
190 - 3225 Derning Way
Middleton
53562
Wisconsin
UNITED STATES