Health product recall

EMU40 White and EMU40EX (2018-05-16)

Starting date:
May 16, 2018
Posting date:
June 14, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67028

Affected products

A. EMU40 White

B. EMU40EX

Reason

Due to degradation of an internal electronic component over time the EMU40/EMU40EX breakout box may fail. This rare case of component degradation occurs over time and occurs when the breakout box exceeds its expected lifespan of 7 years.

Affected products

A. EMU40 White

Lot or serial number

All lots.

Model or catalog number

002933

Companies
Manufacturer

Natus Medical Incorporated DBA Excel-Tech Ltd (XLTEK)

2568 Bristol Circle

Oakville

L6H 5S1

CANADA



B. EMU40EX

Lot or serial number

All lots.

Model or catalog number

10406

Companies
Manufacturer

Natus Medical Incorporated DBA Excel-Tech Ltd (XLTEK)

2568 Bristol Circle

Oakville

L6H 5S1

CANADA