Health product recall

EMS Power-LOAD (2018-03-12)

Starting date:
March 12, 2018
Posting date:
April 13, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type I
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66410

Affected products

EMS Power-LOAD

Reason

It was reported via three complaints that after installation of the power-LOAD, the floor plate attachment bracket (PN: 6390-101-108) was found to be broken.
After installation of the power-LOAD, the floor plate attachment brackets (PN: 6390-101-108) which allow the power-LOAD anchor to be secured to the power-load floor plate may be found to be broken after removing the anchor. All power-LOAD units manufactured from January 1, 2016 to December 31, 2016, may potentially exhibit this issue.

Affected products

EMS Power-LOAD

Lot or serial number

More than 100 numbers, contact manufacturer.

Model or catalog number

6390-000-000

Companies
Manufacturer
Stryker Medical
3800 E Centre Avenue
Portage
49002
UNITED STATES