Issue:
Medical Devices
Audience:
General Public, Hospitals, Healthcare Professionals
Identification number:
RA-59418

Affected products

A. EMPTY,STERILE I.V.,SOLUTION TRANSFER CONTAINERS

Reason

AutoComp6, AutoComp6- XP and AutoComp6-XPS high speed compounders may not be able to maintain the fill accuracy requirements for the compounded bags as outlined in the user's manual. Therefore, the patient receiving the compounded bag may both receive the correct quantity of individual parenteral nutrition solutions or the correct overall quantity of parenteral nutrition solutions.

Affected products

A. EMPTY,STERILE I.V.,SOLUTION TRANSFER CONTAINERS

Lot or serial number

All lots

Model or catalog number
  • "58740
  • 58741
  • 58742
  • 58750
  • 58753
  • 58754
  • 58781
  • 58785
  • 66005
  • 66010
  • 66015
  • 66020
  • 66025
  • 66030
  • 66035
  • 66310
  • 66315
  • 66325
  • 66330
  • 66710
  • 66715"
Companies
Manufacturer

THE METRIX COMPANY

4400 CHAVENELLE RD

DUBUQUE

52002

Iowa

UNITED STATES