Emprint Percutaneous Antenna with Thermosphere Technology Large Spherical Ablation Zone (2019-03-28)
- Starting date:
- March 28, 2019
- Posting date:
- May 3, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-69812
Last updated: 2019-05-03
Affected Products
Emprint Percutaneous Antenna with Thermosphere Technology Large Spherical Ablation Zone
Reason
Medtronic is advising customers that we will be updating the Instructions for Use (IFUs) to include important information identified through our investigation of detached antenna tips. Tip detachment during a procedure can lead to a retained foreign body, which creates risk for infection.
Affected products
Emprint Percutaneous Antenna with Thermosphere Technology Large Spherical Ablation Zone
Lot or serial number
All lots.
Model or catalog number
CA15L1
CA15L2
CA20L1
CA20L2
CA30L1
CA30L2
Companies
- Manufacturer
-
Covidien LLC
15 Hampshire Street
Mansfield
02048
Massachusetts
UNITED STATES