This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Embryoscope ES-D (2014-04-24)

Starting date:
April 24, 2014
Posting date:
October 27, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-41897

Recalled Products

Embryoscope ES-D 

Reason

When cleaning or disinfecting parts of the Embryoscope time-lapse incubator, the metal clips which are attached to the embryo chamber and support the transparent plate positioned over the culture dish holder may come loose.

Affected products

Embryoscope ES-D 

Lot or serial number

Not applicable 

Model or catalog number

ESD20000-JAP

Companies
Manufacturer

Unisense Fertilitech A/S,

Jens Juuls Vej 20, Viby J,

Viby J 8260

DENMARK