Emblem S-Icd and Latitude Programming Systems
Brand(s)
Last updated
Summary
Product
Emblem S-Icd and Latitude Programming Systems
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
Emblem S-Icd System | Not applicable. | 3200 |
Latitude Programming System | Not applicable. | 3300 |
Issue
In December 2020, Boston Scientific committed to developing a software enhancement that detects and alerts Healthcare Professionals (HCPs) if an Emblem S-Icd exhibits hydrogen-induced accelerated battery depletion. This enhanced software allows clinicians to identify Emblem S-Icds exhibiting hydrogen-induced accelerated battery depletion sooner. This enhanced software is now available, and sales professionals are upgrading the programmers.
Recall start date: July 7, 2022
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies
Boston Scientific Corporation
4100 Hamline Avenue North, St. Paul, Minnesota, United States, 55112-5798
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-64393