Health product recall

Ellex Solo-A (2020-04-03)



Starting date:
April 3, 2020
Posting date:
May 1, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-72941


Last updated:

Affected Products

Ellex Solo-A

Reason

Defect in the joystick wiring. There is the potential that the insulation of the wire can become compromised, potentially causing the exposed conductor to come into contact, with the grounded metal housing. As an outcome when the system is in the "ready" mode, an unintended laser emission may result.

Affected products

Ellex Solo-A

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • LT5106-S
Companies
Manufacturer

ELLEX MEDICAL PTY LTD

3-4 Second Ave.

Mawson Lakes

5095

AUSTRALIA