Elekta Unity (2020-05-06)
- Starting date:
- May 6, 2020
- Posting date:
- June 5, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-73249
Last updated: 2020-06-05
Affected Products
Elekta Unity
Reason
There is a potential risk where legionella bacteria may be present in the Unity machine room, specifically the heat exchanger condensation collection tray.This risk is only significant to service users working directly on the condensation collection tray and immediate area. Clinical users and patients have no exposure to this potential hazard.
Affected products
Elekta Unity
Lot or serial number
- S/N: 600012
- S/N: 600020
Model or catalog number
- 1536549
Companies
- Manufacturer
-
Elekta Limited
Linac House, Fleming Way
Crawley, Sussex
RH10 9RR
UNITED KINGDOM