Health product recall

Electrosurgical Generator, ESG-410

Last updated

Summary

Product
Electrosurgical Generator, ESG-410
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
Electrosurgical Generator, ESG-410 More than 10 numbers, contact manufacturer. WA91327U

Issue

Olympus is conducting this medical device recall after identifying an increase in complaints indicating an occurrence of system error e1226 "could not write to audio buffer" during procedures using the ESG-410 (Sep). This error is a high priority alarm which triggers an automatic reboot of the device, which may lead to a delayed or interrupted procedure until the reboot is completed. Rebooting is a safety measure to put the generator back into safe operation. After the reboot, the procedure can be continued. Olympus has developed a software update that prevents this system error e1226 from unnecessarily occurring.
Olympus has received six (6) complaints for the ESG-410 (sep) generator with the occurrence of system error e1226. No serious injuries or reportable malfunctions have been reported in relation to this issue.

Recall start date: October 1, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies
Olympus Winter & Ibe Gmbh
Kuehnstrasse 61, Hamburg, Hamburg, Germany, 22045
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76340

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe