Elecsys® Troponin T Hs
Brand(s)
Last updated
Summary
Product
Elecsys® Troponin T Hs
Issue
Medical devices - Performance issue
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Industry
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
Elecsys® Troponin T Hs | 61491501 64100801 63986301 64241701 | 08469717190 08469873190 |
Elecsys® Troponin T Hs Stat | 59638102 56661101 59638101 63480901 56661102 56661103 | 08469814190 |
Issue
During internal studies with the Elecsys® Troponin T hs (high sensitive) and Elecsys® Troponin T hs Stat assays, it was confirmed that for certain k2/k3 EDTA primary tubes, TnT hs results are elevated compared to serum samples when measured from the primary tube after processing the sample according to the tube manufacturers' instructions. This issue does not affect lithium heparin plasma or serum sample tubes.
Recall start date: January 16, 2023
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies
Roche Diagnostics GmbH
Sandhoferstrasse 116, Mannheim, Germany, 68305
Published by
Health Canada
Audience
Healthcare
Industry
Recall class
Type II
Identification number
RA-72632
Get notified
Receive notifications for new and updated recalls and alerts by category.