Health product recall

Elecsys Calcitonin (2019-10-07)

Starting date:
October 7, 2019
Posting date:
November 1, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-71467


Last updated:

Affected Products

Elecsys Calcitonin

Reason

The manufacturer performed drug interference testing and confirmed that an influence on recovery was found for itraconazole on Elecsys Calcitonin assays. The method sheets will be revised to include a disclaimer in the limitation-interference section.

Affected products

Elecsys Calcitonin

Lot or serial number

All lots.

Model or catalog number

6445853190

7027044190

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY