Health product recall

EGM - Essential Gas Module

Last updated

Summary

Product
EGM - Essential Gas Module
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare
Industry

Affected products

Affected products Lot or serial number Model or catalogue number
EGM - Essential Gas Module Not applicable. 866173

Issue

Philips has identified a potential safety issue for the 866173 IntelliVue G7m anesthesia gas modules, which could have a component issue that may trigger an interruption of the gas measurement. This issue concerns the G7m's sample gas pump which could stop working, producing a technical, inop (inoperable) alarm. It was anticipated that the expected operating lifetime of the pump was at least 10000 hours. However, based on testing, the average operating lifetime was found to be 4500 hours. The pump may fail prematurely when there is friction caused by an abrasion.

Recall restart date: February 14, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies
Philips Medizin Systeme Boblingen Gmbh
Hewlett-Packard Strasse 2, Boblingen, Germany, 71034
Published by
Health Canada
Audience
Healthcare
Industry
Recall class
Type II
Identification number
RA-76646