EGM - Essential Gas Module
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
EGM - Essential Gas Module |
Not applicable. |
866173 |
Issue
Philips has identified a potential safety issue for the 866173 IntelliVue G7m anesthesia gas modules, which could have a component issue that may trigger an interruption of the gas measurement. This issue concerns the G7m's sample gas pump which could stop working, producing a technical, inop (inoperable) alarm. It was anticipated that the expected operating lifetime of the pump was at least 10000 hours. However, based on testing, the average operating lifetime was found to be 4500 hours. The pump may fail prematurely when there is friction caused by an abrasion.
Recall restart date: February 14, 2023
Additional information
Details
| Philips Medizin Systeme Boblingen Gmbh |
| Hewlett-Packard Strasse 2, Boblingen, Germany, 71034 |
Get notified
Receive emails about new and updated recall and safety alerts.