Health product recall

Eea Autosuture Circular Stapler With Dst Series

Brand(s)
Last updated

Summary

Product
Eea Autosuture Circular Stapler With Dst Series
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Eea Autosuture Circular Stapler With Dst Series

More than 10 numbers, contact manufacturer.

EEA25
EEAXL25
EEA2535
EEAXL2535

Issue

Notification to inform customers regarding potential for the staple guide to not be securely attached to the instrument. A staple guide not attached to the instrument could cause the component to disengage and if disengaged, could allow the device to transect tissue without forming staples. This could result in delay of treatment, extended hospital stay, unspecified tissue injury, unintended radiation exposure, unexpected medical intervention, foreign body in patient, failure to anastomose, and hemorrhage.

Recall start date: May 16, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies

Covidien LLC

15 Hampshire Street, Mansfield, Massachusetts, United States, 02048

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-64233

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe