Eclipse Mobility Scooters
Brand(s)
Last updated
Summary
Product
Eclipse Mobility Scooters
Issue
Medical devices - Performance issue
What to do
Instructions provided to dealers on how to do corrections for customers.
Audience
Healthcare
Industry
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Eclipse Mobility Scooters |
Contact the manufacturer. |
S148 |
Issue
There is a possibility that a small portion of one seat pin batch under the seat base of the affected mobility scooter(s) was not forged in the right conditions. This may result in insufficient strength and may cause a breakage. In the event of an accident where the seat pin breaks, the occupant may fall and cause an injury.
Recall start date: May 18, 2021
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Physical medicine
Companies
Eclipse Medical Inc.
725 Eaton Way, Delta, British Columbia, Canada, V3M 6S5
Published by
Health Canada
Audience
Healthcare
Industry
Recall class
Type II
Identification number
RA-73849