This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
Eclipse
- Starting date:
- September 30, 2005
- Posting date:
- November 8, 2005
- Type of communication:
- Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Identification number:
- RA-29428
Reason
If the modification made to a plan affects the planning field's dose contribution to the primary reference point the treatment field can change without warning which could result in an overdose.
Affected products
Eclipse
Lot or serial number
H480001-H484301
Model or catalog number
N/A
Companies
- Manufacturer
-
Varian Medical System, Inc.