This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
Eclipse
- Starting date:
- September 30, 2005
- Posting date:
- November 28, 2005
- Type of communication:
- Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Identification number:
- RA-29458
Reason
Dose calculations for ARC fields using the pencil beam convolution with the generalized Batho power law or modified Batho heterogeneity correction methods can be wrong if the beam data contains XXX.LIH file.
Affected products
Eclipse
Lot or serial number
H480001-H484301
Model or catalog number
N/A
Companies
- Manufacturer
-
Varian Medical System, Inc.