Health product recall

Echo Bi-Metric Collarless Porous Stem, Size 7 (2017-12-28)

Starting date:
December 28, 2017
Posting date:
April 24, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66578

Affected products

Echo Bi-Metric Collarless Porous Stem, Size 7

Reason

Zimmer Biomet is conducting a lot specific recall for the Echo Bi-Metric Collarless Porous Stem (size 7). In 2008, the Echo Bi-Metric Hip Stems underwent a design change to reduce the lateral shoulder of the device in order to make insertion of the device easier during surgery. It was later determined that item 195207 has the potential to fracture based on insertion testing.

Affected products

Echo Bi-Metric Collarless Porous Stem, Size 7

Lot or serial number
  • 485280
  • 512500
  • 762990
  • 903550
Model or catalog number
  • 192507
Companies
Manufacturer
Biomet Orthopedics
56 East Bell Drive
Warsaw
46581
UNITED STATES