Health product recall

Easystand Bantam

Last updated

Summary

Product
Easystand Bantam
Issue
Medical devices - Performance issue
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Easystand Bantam Small And Easystand Bantam Extra Small

V2005695011
V2005693012

PT50002-1 WITH NON-SUPINE OPTION PT50232

PT50001-1

Issue

Discovered through customer complaint on 9/23/22 and 9/30/22. This issue may not effectively lock the front footrest carrier bracket to the stud on the main frame locking out the supination of that bracket. This results in the unit going into the supine position.

Recall start date: October 4, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Physical medicine
Companies

Altimate Medical Inc.

262 W 1st Street, Morton, Minnesota, United States, 56270

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-72189

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe