Affected products

Affected products

Lot or serial number

Model or catalogue number

Easystand Bantam Small And Easystand Bantam Extra Small

V2005695011
V2005693012

PT50002-1 WITH NON-SUPINE OPTION PT50232

PT50001-1

Issue

Discovered through customer complaint on 9/23/22 and 9/30/22. This issue may not effectively lock the front footrest carrier bracket to the stud on the main frame locking out the supination of that bracket. This results in the unit going into the supine position.