Easy Swivel Carry Bar
- Starting date:
- October 30, 2012
- Posting date:
- November 26, 2012
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public (GP), Healthcare Professionals, Hospitals
- Identification number:
- RA-15911
Recalled Products
- Easy Swivel Carry Bar
Reason
The manufacturer received a report of a lift inspection in the UK where a lift was discovered with a strap that was showing indications of wear inside the easy swivel carry bar. The root cause is a non-conforming insert in the carry bar. The 360760 easy swivel carry bars are impacted.
Affected products
A. Easy Swivel Carry Bar
Lot or serial number
Not applicable
Model or catalog number
360760
Companies
- Manufacturer
- Prism Medical USA