This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Easy Swivel Carry Bar

Starting date:
October 30, 2012
Posting date:
November 26, 2012
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public (GP), Healthcare Professionals, Hospitals
Identification number:
RA-15911

Recalled Products

  1. Easy Swivel Carry Bar

Reason

The manufacturer received a report of a lift inspection in the UK where a lift was discovered with a strap that was showing indications of wear inside the easy swivel carry bar.  The root cause is a non-conforming insert in the carry bar.  The 360760 easy swivel carry bars are impacted.

Affected products

A. Easy Swivel Carry Bar

Lot or serial number

Not applicable

Model or catalog number

360760

Companies
Manufacturer

Prism Medical USA