Health product recall

Easy Moving Plus B Mobile X-Ray Unit With Digital Detector

Last updated

Summary

Product
Easy Moving Plus B Mobile X-Ray Unit With Digital Detector
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
Easy Moving Plus B Mobile X-Ray Unit With Digital Detector G71467 G77408 G73624 G82557 G79899 G82499 SM-40HF-B-D-VIR

Issue

Notification to inform customers regarding potential safety issue related to the steel cable installed inside the mobile column which supports the weight of telescopic arm with the tube head. The potential risk consists of an interference which could cause that the safety system which blocks the arm of the cable is cut off does not work properly. The interference occurs between the cable and a metal plate of the arm's carriage due to shape of the cable's crimped terminals. Scope of the safety recall is to review the installed base according to the manufacturer's instructions and install the kit immediately if the cable could interfere with the parachute system.

Recall start date: June 5, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
Sociedad Espanola De Electromedicina Y Calidad, S.A. (Sedecal)
Poligono Industrial Rio De Janeiro, Algete, Madrid, Spain, 28110
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75707

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe