Easy Moving Plus B Mobile X-Ray Unit With Digital Detector
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Easy Moving Plus B Mobile X-Ray Unit With Digital Detector |
G71467 |
SM-40HF-B-D-VIR |
Issue
Notification to inform customers regarding potential safety issue related to the steel cable installed inside the mobile column which supports the weight of telescopic arm with the tube head. The potential risk consists of an interference which could cause that the safety system which blocks the arm of the cable is cut off does not work properly. The interference occurs between the cable and a metal plate of the arm's carriage due to shape of the cable's crimped terminals. Scope of the safety recall is to review the installed base according to the manufacturer's instructions and install the kit immediately if the cable could interfere with the parachute system.
Recall start date: June 5, 2024
Additional information
Details
| Sociedad Espanola De Electromedicina Y Calidad, S.A. (Sedecal) |
| Poligono Industrial Rio De Janeiro, Algete, Madrid, Spain, 28110 |
Get notified
Receive emails about new and updated recall and safety alerts.