DxC 500i Clinical Analyzers
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
DxC 500i Clinical Analyzer, DxC 500i Au Chemistry Module |
software 1.3.3 or 1.3.4 |
C63522 |
DxC 500i Clinical Analyzer, DxC 500i Access 2 Immunoassay Module |
software 1.3.3 or 1.3.4 |
C13252 |
Issue
Beckman Coulter has determined that on the DxC 500i clinical analyzer, when ordering a clinical chemistry (CC) combination test simultaneously with an immunoassay (IA) test, if the IA test is processed between the CC constituent tests, the sample status of the CC test will remain "in progress" with no error events.
Recall start date: August 4, 2025
Additional information
Details
| Beckman Coulter, Inc. |