DxC 500 AU Clinical Chemistry Analyzer
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
DxC 500 AU Clinical Chemistry Analyzer |
2024070118 |
C63520 |
Issue
Beckman Coulter has determined that the system cannot recalculate a calculated result when results are rerun within the calculated test. The system will report only the calculated result using the initial constituent tests' results to the operator and lis/middleware. A patient may receive unnecessary treatment or testing; or patient treatment may be delayed.