Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

DxC 500 AU Clinical Chemistry Analyzer

2024070118

C63520
C63519

Issue

Beckman Coulter has determined that the system cannot recalculate a calculated result when results are rerun within the calculated test. The system will report only the calculated result using the initial constituent tests' results to the operator and lis/middleware. A patient may receive unnecessary treatment or testing; or patient treatment may be delayed.

Recall Start Date: February 13, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies

Beckman Coulter, Inc.

250 S. Kraemer Blvd., Brea, California, United States, 92821

Published by
Health Canada
Audience
Healthcare
Recall class
Type II