DUREX Real Feel Extra Lubricated 10ct condoms (2018-07-30)
- Starting date:
- July 30, 2018
- Posting date:
- August 17, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-68726
Affected products
DUREX Real Feel Extra Lubricated 10ct condoms
Reason
Following observation of variable burst pressure performance of polyisoprene condoms manufactured during 2018; Reckitt Benckise conducted an internal investigation and identified that one batch of Non-Latex Durex RealFeel Extra Lubricated 10ct Condoms, UPC: 0 67981 98715 7, Batch: 1000443254, Expiry Date: 2023 - 02 marketed in Canada between June 2018 to July 2018 is not expected to meet the internal burst pressure specification of 1.0 kPa at the end of 5 year shelf-life. Consequently, Reckitt Benckiser (Canada) Inc. as a cautious approach is initiating a consumer level recall.
Affected products
DUREX Real Feel Extra Lubricated 10ct condoms
Lot or serial number
1000443254
Model or catalog number
98715
Companies
- Manufacturer
-
Reckitt Benckiser (Canada) Inc
1680 Tech Avenue, Unit 2
Mississauga, Ontario
L4W 5S9
CANADA