This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Duradiagnost

Starting date:
May 26, 2017
Posting date:
June 13, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-63612

Reason

Philips discovered that some welding joints on the vertical carriage brackets of the tube stand may crack. As a result, the tube arm assembly may exhibit uncharacteristic sideways movement, with the tube becoming off center. If all of the welding joints cracked, the tube arm assembly could fall down.

Affected products

Duradiagnost

Lot or serial number

All lots

Model or catalog number

9897-010-02021

Companies

Manufacturer

PHILIPS HEALTHCARE (SUZHOU) CO., LTD

NO. 258, ZHONG YUAN ROAD

SUZHOU

215024

CHINA