This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
Duradiagnost
- Starting date:
- May 26, 2017
- Posting date:
- June 13, 2017
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-63612
Reason
Philips discovered that some welding joints on the vertical carriage brackets of the tube stand may crack. As a result, the tube arm assembly may exhibit uncharacteristic sideways movement, with the tube becoming off center. If all of the welding joints cracked, the tube arm assembly could fall down.
Affected products
Duradiagnost
Lot or serial number
All lots
Model or catalog number
9897-010-02021
Companies
- Manufacturer
-
PHILIPS HEALTHCARE (SUZHOU) CO., LTD
NO. 258, ZHONG YUAN ROAD
SUZHOU
215024
CHINA