This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

DUAL HEMO MCABLE PODS (2017-10-02)

Starting date:
October 2, 2017
Posting date:
November 7, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type I
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-65040

Affected products

DUAL HEMO MCABLE PODS

Reason

Draeger Medical Systems inc.  has determined through internal testing that Dual Hemo MCable Pods with Revision Index (RI) 15 and 16, which are used with the Draeger Infinity Acute Care System, may permit liquid ingress inconsistent with their IPX4 rating, which could potentially result in an incorrect measured value of invasive blood pressure.

Affected products

DUAL HEMO MCABLE PODS

Lot or serial number

More than 100 numbers, contact manufacturer

Model or catalog number

MS20783

Companies
Manufacturer

Draeger Medical Infant Care, Inc.

3135 QUARRY RD,

TELFORD

18969

Pennsylvania

UNITED STATES