DUAL HEMO MCABLE PODS (2017-10-02)

Starting date:
October 2, 2017
Posting date:
November 7, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type I
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-65040

Affected products

DUAL HEMO MCABLE PODS

Reason

Draeger Medical Systems inc.  has determined through internal testing that Dual Hemo MCable Pods with Revision Index (RI) 15 and 16, which are used with the Draeger Infinity Acute Care System, may permit liquid ingress inconsistent with their IPX4 rating, which could potentially result in an incorrect measured value of invasive blood pressure.

Affected products

DUAL HEMO MCABLE PODS

Lot or serial number

More than 100 numbers, contact manufacturer

Model or catalog number

MS20783