DUAL HEMO MCABLE PODS (2017-10-02)
- Starting date:
- October 2, 2017
- Posting date:
- November 7, 2017
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type I
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-65040
Affected products
DUAL HEMO MCABLE PODS
Reason
Draeger Medical Systems inc. has determined through internal testing that Dual Hemo MCable Pods with Revision Index (RI) 15 and 16, which are used with the Draeger Infinity Acute Care System, may permit liquid ingress inconsistent with their IPX4 rating, which could potentially result in an incorrect measured value of invasive blood pressure.
Affected products
DUAL HEMO MCABLE PODS
Lot or serial number
More than 100 numbers, contact manufacturer
Model or catalog number
MS20783
Companies
- Manufacturer
-
Draeger Medical Infant Care, Inc.
3135 QUARRY RD,
TELFORD
18969
Pennsylvania
UNITED STATES